255
Specifications
Performance Specifications
Analog output For pressure monitoring, 2mV/bar, one output
Communications Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN optional
Safety and maintenance Extensive diagnostics, error detection and display (through control module
and AgilentChemSt ation), leak detection, safe leak handling, leak output
signal for shutdown of pumping system. Low voltages in major
maintenance areas.
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of seal wear and volume of pumped mobile phase with
user-settable limits and feedback messages. Electronic records of
maintenance and errors.
Housing All materials recyclable.
Table 44 Performance Specification Agilent1100 Series Quaternary Pump