Troubleshooting and Test Functions

Temperature Limit Exceeded

Temperature Limit Exceeded

 

The temperature of one of the motor-drive circuits is too high.

 

The processor continually monitors the temperature of the drive circuits on

 

the LPM board. If excessive current is being drawn for long periods, the

 

temperature of the circuits increases. If the temperature exceeds the upper

 

limit of 95 ºC, the error message is generated.

Probable Causes

High friction (partial mechanical blockage) in the pump drive assembly.

 

Partial blockage of the flowpath in front of the damper.

 

Defective pump drive assembly.

 

Defective LPM board.

Suggested Actions

Ensure the capillaries and frits between the pump head and damper inlet

 

are free from blockage.

 

Ensure the outlet valve is not blocked.

 

Remove the pump head assembly. Ensure there is no mechanical blockage

 

of the pump head assembly or pump drive assembly.

 

Exchange the pump drive assembly.

 

Exchange the LPM board.