Agilent Technologies 1100 Series manual Exchanging the Active Inlet Valve

Models: 1100 Series

1 272
Download 272 pages 29.99 Kb
Page 103
Image 103

Repairing the Pump

Exchanging the Active Inlet Valve Cartridge or the Active Inlet Valve

 

Exchanging the Active Inlet Valve

 

Cartridge or the Active Inlet Valve

 

 

When required

If internally leaking (backflow)

Tools required

Wrench 14 mm

 

Pair of Tweezers

Material

Active inlet valve G1312-60010 (complete assembly)

 

Valve cartridge 5062-8562

 

 

Removing the Active Inlet Valve

1Unplug the active inlet valve cable from the connector.

2Disconnect the solvent inlet tube at the inlet valve.

3Using a 14 mm wrench loosen the active inlet valve and remove the valve from pump head.

Exchanging the Valve Cartridge

1Using a pair of tweezers remove the valve cartridge from the actuator assembly.

2Before inserting the new valve cartridge clean the area in the actuator assembly. Fill a syringe with alcohol and flush the cartridge area thouroughly.

3Insert a new valve cartridge into the actuator assembly. Make sure the valve cartridge is fully inserted into the actuator assembly.

103

Page 103
Image 103
Agilent Technologies 1100 Series manual Exchanging the Active Inlet Valve, Cartridge or the Active Inlet Valve

1100 Series specifications

Agilent Technologies 1100 Series is a highly regarded lineup of high-performance liquid chromatography (HPLC) systems designed to meet the demanding needs of laboratories in various fields including pharmaceuticals, environmental testing, and food safety. Known for its versatility, reliability, and advanced technology, the 1100 Series has established itself as a benchmark for analytical performance.

One of the standout features of the Agilent 1100 Series is its modular architecture. This design allows users to configure the system according to specific analytical requirements. The components such as pumps, detectors, and autosamplers can be mixed and matched to create a tailored solution, ensuring maximum efficiency and optimal performance. This modularity also facilitates easy upgrades and maintenance, significantly extending the life cycle of the instrument.

At the heart of the 1100 Series is its advanced pumping technology. The system utilizes precise, high-pressure pumps that ensure a consistent flow rate and excellent reproducibility, essential for accurate quantification of analytes. The binary and quaternary pump options provide flexibility in method development, enabling users to perform gradient elution with ease.

The Agilent 1100 Series also boasts a range of detectors, including UV/VIS, fluorescence, and refractive index detectors. Each detector is designed to enhance sensitivity and specificity, enabling the detection of a wide array of compounds. The UV/VIS detector, in particular, is known for its reliable performance and user-friendly interface, which simplifies method setup and data analysis.

Another critical aspect of the 1100 Series is its autosampler. The system's autosamplers provide high throughput and automatic sample handling, which significantly reduces human error and increases productivity. The intelligent design allows for temperature control and reliable sample injection, ensuring optimal results.

In terms of software, the 1100 Series is supported by Agilent’s OpenLAB software, which offers an intuitive platform for method development, data acquisition, and analysis. This software integrates seamlessly with the HPLC hardware, making it easier for scientists to manage workflows and maintain compliance with regulatory requirements.

Overall, the Agilent Technologies 1100 Series represents a significant advancement in liquid chromatography, bringing together ideal characteristics such as modularity, precision, versatility, and sophistication. Its commitment to innovation and quality makes it a preferred choice for laboratories seeking reliable analytical solutions.