Repairing the Pump

Exchanging the Low Pressure Pump Main Board (LPM Board)

Exchanging the Low Pressure Pump Main

Board (LPM Board)

When required

Board defective

Tools required

Wrench 14 mm

 

Wrench 7 mm

 

Wrench 5 mm

Parts required

LPM Board, G1311-66520, exchange part number G1311-69520

 

 

1Turn off the pump, disconnect all cables and remove the pump from the stack.

2Remove the top cover and foam (see “Removing the Top Cover and Foam” on page 128).

3Disconnect all connectors from the main board.

N O T E

When removing connectors, counterhold on connector J3 with one hand.

Figure 20

LPM Board Connectors

 

 

J4

J1

 

 

S1

 

J5

J2

J3

 

 

J6

 

 

 

 

 

J11

 

J9

 

 

 

 

 

J19

J16

J17

J23

 

 

 

 

 

 

 

 

 

 

J26

 

J30

 

 

 

J24

J25

 

 

 

 

 

 

 

 

 

131

Page 131
Image 131
Agilent Technologies 1100 Series manual Exchanging the Low Pressure Pump Main Board LPM Board, LPM Board Connectors

1100 Series specifications

Agilent Technologies 1100 Series is a highly regarded lineup of high-performance liquid chromatography (HPLC) systems designed to meet the demanding needs of laboratories in various fields including pharmaceuticals, environmental testing, and food safety. Known for its versatility, reliability, and advanced technology, the 1100 Series has established itself as a benchmark for analytical performance.

One of the standout features of the Agilent 1100 Series is its modular architecture. This design allows users to configure the system according to specific analytical requirements. The components such as pumps, detectors, and autosamplers can be mixed and matched to create a tailored solution, ensuring maximum efficiency and optimal performance. This modularity also facilitates easy upgrades and maintenance, significantly extending the life cycle of the instrument.

At the heart of the 1100 Series is its advanced pumping technology. The system utilizes precise, high-pressure pumps that ensure a consistent flow rate and excellent reproducibility, essential for accurate quantification of analytes. The binary and quaternary pump options provide flexibility in method development, enabling users to perform gradient elution with ease.

The Agilent 1100 Series also boasts a range of detectors, including UV/VIS, fluorescence, and refractive index detectors. Each detector is designed to enhance sensitivity and specificity, enabling the detection of a wide array of compounds. The UV/VIS detector, in particular, is known for its reliable performance and user-friendly interface, which simplifies method setup and data analysis.

Another critical aspect of the 1100 Series is its autosampler. The system's autosamplers provide high throughput and automatic sample handling, which significantly reduces human error and increases productivity. The intelligent design allows for temperature control and reliable sample injection, ensuring optimal results.

In terms of software, the 1100 Series is supported by Agilent’s OpenLAB software, which offers an intuitive platform for method development, data acquisition, and analysis. This software integrates seamlessly with the HPLC hardware, making it easier for scientists to manage workflows and maintain compliance with regulatory requirements.

Overall, the Agilent Technologies 1100 Series represents a significant advancement in liquid chromatography, bringing together ideal characteristics such as modularity, precision, versatility, and sophistication. Its commitment to innovation and quality makes it a preferred choice for laboratories seeking reliable analytical solutions.