227
Performance Tests
V. Polarization Dependent Loss (PDL): Optional
Performance Test for the Agilent 8156A Option 201
Page 6 of 8
Mode l Agilent 8156A Attenuator Option 201 No. _______________ Date_______________
Test
No. Test Description
perfo rmed at _________________nm Minimum
Spec. Result Maximum
Spec. Measurement
Uncertainty
I. Total Insertion Loss Test dB ±0.60dB
typ. <2.5dB
measured at __________________nm
with SM fiber _________ 3.0 dB
II. Linearity/Att. Acc. ±0.05dB
Attenuation
Setting:
0dB REF
1dB 0.9dB _________ 1.1dB
2dB 1.9dB _________ 2.1dB
3dB 2.9dB _________ 3.1dB
4dB 3.9dB _________ 4.1dB
5dB 4.9dB _________ 5.1dB
6dB 5.9dB _________ 6.1dB
7dB 6.9dB _________ 7.1dB
8dB 7.9dB _________ 8.1dB
9dB 8.9dB _________ 9.1dB
10dB 9.9dB _________ 10.1dB