113
Remote Commands
OUTPut Commands

:OUTPut:POWer?

Syntax :OUTPut:POWer? [<wsp> MIN|DEF|MAX]
Description The query returns the current through-power,
in dBm.
ThroughPower(dBm) = ThroughPowerBase(dBm) + Attfilter@Base(dB)-Att
filter(dB)
By sending MIN, DEF, or MAX with the query
the minimum, default or maximum value
possible for the through-p ower is returned.
Example OUTPUT 728;":OUTP:POW 32.15"
OUTPUT 728;":OUTP:POW?"
ENTER 728;A$

:OUTPut:[:STATe]

Syntax :OUTPut[:STATe] <wsp> OFF|ON|0|1
Description This command sets the state of the output
shutter, that is, whether it is open or closed.
OFF or 0 closes the shutter, and no power gets
through. ON or 1 opens the shutter, and power
gets through.

:OUTPut[:STATe]?

Syntax :OUTPut[:STATe]?
Description The query returns whether t he ou tput s hutter is
open or closed.
0 indicates the shutter is closed (no power is
getting through). 1 indicates that the shutter is
open (power is getting through).
Example OUTPUT 728;":OUTP ON"
OUTPUT 728;":OUTP?"
ENTER 728;A$