GPIB Messages

-314 Save/recall memory lost

The nonvolatile data saved by the *SAV command has been lost.

-315 Configuration memory lost

The nonvolatile configuration data saved by the instrument has been lost.

-330 Self-test failed

Further information about the self-test failure is available by using *TST?.

-350 Queue overflow

The error queue has overflown. This error is written to the last position in the queue, no further errors are recorded.

Query Errors

These are error messages in the range -400 to -499. They indicate that an error has been detected by the output queue control.

An device-specific error is signaled by the query error bit (bit 2) in the event status register.

-300 Query error

This indicates that a query error has occurred. No more specific information is available.

-410 Query INTERRUPTED

A condition occurred which interrupted the transmission of the response to a query (for example, a query followed by a DAB or a GET before the response was completely sent).

-420 Query UNTERMINATED

A condition occurred that interrupted the reception of a query (for example, the instrument was addressed to talk and an incomplete program message was received).