112
Remote Commands
OUTPut Commands
Example OUTPUT 728;":INP:ATT?"
ENTER 728; Att
OUTPUT 728;":INP:OFFS?"
ENTER 728; Cal
Newcal = Basepow - Att + Cal
OUTPUT 728;":INP:OFFS ";Newcal
OUTPUT 728;":OUTP:APM ON"
OUTPUT 728;":OUTP:APM?"
ENTER 728;A$

:OUTPut:POWer

Syntax :OUTPut:POWer <wsp>
<value>[DBM]|MIN|DEF|MAX
Description This command sets the through-power for the
instrument. The through-power is used to set
the attenuation of the filter.
Attfilter(dB) = ThroughPowerBase(dBm) - ThroughPower(dB m) + At tfilter@Base(dB)
You set the through-power by sending a value
(default units are dBm), or by sending MIN,
DEF or MAX, which specify the minimum,
default and maximum values for the through-
power.
The maximum value and the default value are
those values for which Attenuationfilter =0dB.
The minimum value is that value for which
Attenuationfilter is at its greatest. For example,
if you have set INP:ATT 10 and INP:OFFS
2 and then switched UTP:APM ON, then the
throug h po wer is set t o 12 dB m. The max i mum
through power, and the default through power,
in this case is 22dBm. The minimum throug h
power in this case is -38dBm.