180
Performance Tests
Performance Test
Figure D-2 Tot al Insertion Los s Test Setup 1, Options 201, 221
Figure D-3 Total Insertion Loss Test Setup 1, Option 350
3. On the DUT, press and hold ATT to reset the attenuation to
minimum (any attenuation shown on the display is d ue to the
calibration factor).
4. Zero the Power-meter and select Autorange. Display [dB]
5. Enable the laser source and set Display to Reference on the
powe r meter.
6. Connect the equipment as shown in the appropriate Total
Insertion Loss Test Setup 2.