170
Specifications
Specifications
λ: Entering of wavelength for automatic correction of attenuatio n
using typical correction values.
Cal: Offset factor to adjust the attenuation factor on the display
within ±99.999dB range.
DispCal: Sets attenuation value on the display to 0.000dB.
Swp: Manual or automatic up or down attenuation sweep. Start,
stop, step size and dwell time (not for manual sweep) can be
entered.
Back Refl : Desired return loss (back refl ection level) can be
entered. Requires Agilent 81000BR back reflector, or Option 203.
Enb/Dis: Optical signal path interrupted wit h shutter (>80dB
isolation).
Store/Recall: 9 user-selectable parameter settings may be stored
and recalled. Recall of default setting.

General

Recalibration period: 1 year.
Warm-up time: 45 Minutes. Not required if previously stored
within operating temperatur e rang e.
GPIB Capability: All modes and parameters can be
programmed, SCPI comma n d s et, 81 57A comp atibility mode.
GPIB Interface Function Code: SH1, AH1, T6, L4, SR1,
RL 1, PP 0, DC2, DT0, C0

Envir onmental

Storage temperature: -40 to +70°C
Operating temperature: 0 to +55°C
Humidity: <9 5% R.H . (to 40°C)
Altitude: to 10,000 feet