53
Making an A ttenuation Swe ep
The Manual Sweep

Resetting the Parameters

To reset any of the sweep parameters, press and hold SWP until the
value reset s (thi s take s appr oxi mat el y two second s).
START and STOP reset so that the filter attenuation (inside the
instrument) is zero, th at is
Start = Cal
or
Stop = Cal
See Entering a Calibration Factor on page 39 for informa tion
about setting the calibration factor, Cal.
STEP resets to zero.

Executing the Manual Sweep

If you have just set up your sweep, then you only need to press
EXEC to run the application.
If you have already set up the sweep, and are currently operating
the instrument as an attenuator,
1. Press SWP, and then,
2. Press EXEC.
Figure 3-6 Running the Ma nual Sweep
If there is something wrong with a parameter ( if STEP is zero, for
example), this parameter is shown on the disp lay fo r editin g. Edit
the parameter, and press EXEC again.