Installation

AC Line Power Supply Requirements

Before switching on the instrument, the protective earth terminal of the instrument must be connected to a protective conductor. You can do this by using the power cord supplied with the instrument.

It is prohibited to interrupt the protective earth connection intentionally.

The following work should be carried out by a qualified electrician. All local electrical codes must be strictly observed. If the plug on the cable does not fit the power outlet, or if the cable is to be attached to a terminal block, cut the cable at the plug end and rewire it.

The color coding used in the cable depends on the cable supplied. If you are connecting a new plug, it should meet the local safety requirements and include the following features:

Adequate load-carrying capacity (see table of specifications).

Ground connection.

Cable clamp.

The AC power requirements are summarized on the rear panel of the instrument.

Figure A-2

Rear Panel Markings

145

Page 145
Image 145
Agilent Technologies 8156A manual Figure A-2 Rear Panel Markings

8156A specifications

Agilent Technologies 8156A is a high-performance optical source offering advanced capabilities for testing and characterizing optical systems. Specifically designed for applications in telecommunications and data communications, the 8156A stands out due to its precision, versatility, and reliability, making it an essential tool for engineers and researchers in the telecommunications industry.

One of the main features of the 8156A is its ability to generate a stable, accurate, and tunable optical signal. This optical source operates across a wide wavelength range, typically from 1260 nm to 1675 nm, covering key regions used in fiber optics. The tunable laser module is particularly beneficial for applications requiring the testing of fiber optic components, systems, and networks, enabling users to specify any wavelength within this range.

The device employs advanced technologies, including distributed feedback (DFB) laser technology, offering low noise and high output power while ensuring minimal phase and frequency jitter. This results in precise measurements that are crucial for evaluating the performance of optical devices. The 8156A also features a built-in optical power meter, allowing for seamless integration and real-time monitoring of optical power levels during testing.

Another significant characteristic of the 8156A is its user-friendly interface, which includes easy-to-read displays and intuitive controls. This design simplifies operation, allowing users to set up tests quickly and efficiently. Additionally, it provides various output options, including single-mode and multi-mode fiber interface compatibility, expanding its usability across different applications.

The 8156A is equipped to handle numerous protocols and systems, including passive optical networks (PON), optical transport networks (OTN), and traditional wavelength-division multiplexing (WDM) technologies. Its agility in adapting to various protocols enhances its role in research and development settings.

Moreover, the device offers a range of automation features, enabling users to script complex measurement sequences, making it an adaptable solution for both laboratory and field environments. Combined with its robust build quality, the Agilent Technologies 8156A optical source not only ensures reliable operation but also provides a long service life, making it a valuable investment for any optical testing application.