Agilent Technologies manual Performance Test for the Agilent 8156A Option, Result

Models: 8156A

1 289
Download 289 pages 37.75 Kb
Page 203
Image 203

Performance Tests

V. Polarization Dependent Loss (PDL): Optional

Performance Test for the Agilent 8156A Option 100

Page 3 of 8

Model Agilent 8156A Attenuator Option 100

No. _______________ Date_______________

Test

Test Description

Minimum

 

Maximum

Measurement

No.

performed at _________________nm

Spec.

Result

Spec.

Uncertainty

 

 

 

 

 

 

 

I.

Total Insertion Loss Test

 

 

dB

 

± 0.60dB

 

typ. <4.5dB

 

 

 

 

 

 

measured at __________________nm

 

 

 

 

 

 

with singlemode fiber

 

 

_________

5.4 dB

 

II.

Linearity/Att. Acc.

 

 

 

 

± 0.05dB

 

Attenuation

 

 

 

 

 

 

Setting:

 

 

 

 

 

 

0dB

 

 

REF

 

 

 

1dB

0.8dB

_________

1.2dB

 

 

2dB

1.8dB

_________

2.2dB

 

 

3dB

2.8dB

_________

3.2dB

 

 

4dB

3.8dB

_________

4.2dB

 

 

5dB

4.8dB

_________

5.2dB

 

 

6dB

5.8dB

_________

6.2dB

 

 

7dB

6.8dB

_________

7.2dB

 

 

8dB

7.8dB

_________

8.2dB

 

 

9dB

8.8dB

_________

9.2dB

 

 

10dB

9.8dB

_________

10.2dB

 

 

 

 

 

 

 

 

203

Page 203
Image 203
Agilent Technologies manual Performance Test for the Agilent 8156A Option, Result