179
Performance Tests
Performance Test

I. Total Insertion Loss Test

Carry out the following Insertion Loss Test at 1310nm and 1550nm
with single-mod e fiber s usin g the the equipment listed previously.
1. Turn the instruments on an d allow the instruments to war m up.
2. Connect the equipment as shown in the appropriate Total
Insertion Loss Test Setup 1.
Figure D-1 Total Insertion Loss Test Setup 1, Options 100, 101, 121
Specifications Agilent 8156A Typ.
Insertion loss (including both connectors) Option 100 <5.4dB
Option 101 <3.0dB
Option 121 <4.2dB
Option 201 <3.0dB
Option 221 <3.3dB
Option 350 <3.0dB