182
Performance Tests
Performance Test
Figure D-6 Total Insertion Loss Test Setup 2, Option 350
7. Enable the attenuator output and record the power meter reading
(in dB) in the Test Record and check that it is within
specifications.

II. Linearity/Attenuation Accuracy Test

Carry out the following Attenuation Accuracy tests at 1310nm and
1550nm with single -mo de fibe rs using the equipment listed
previously.
Specifications Agilent 8156A
Linearity Option 100 <±0.2dB
Option 101 <±0.1dB
Option 121 <±0.1dB
Option 201 <±0.1dB
Option 221 <±0.1dB
Option 350 <±0.1dB