12-22
Module I/O Specifications
A13 Output
A13 Output
Use extender card 08922-90129.
Use coax jumpers on Plug 1, pins 3, 17 and Plug 3, pin 13.
Power Supplies

Inputs2

OUT_1G_REF only present for RF Generator frequencies from 0 to 291 MHz, used for
frequency translation.
+5 V J2(4) 50 mA
+12 V J2(2) 310 mA
12 V J2(3) 80 mA
+38 V J2(1) 10 mA
GND J1(1,2,4,5,6,7,8,9,10,11,12,13,14,15,16,18,19,20)
J3(1,2,3,4,5,6,7,8,9,10,11,12,14,15,16,17,18,19,20)
From Sum Loop A25 (Change frequency on RF Generator page)
(Coax jumper connection) SGS_500_10000M J1(17)
Freq Range 480 - 1015 MHz
Input Level required 0 dBm ± 2 dB
Spectral Purity required
Harmonics 2nd < -10 dBc 3rd - 5th < -15 dBc
Spurs < -70 dBc
From A15 Reference Assembly
(Coax jumper connection) OUT_1G_REF J1(3)
Input Level 1 dBm ± 3 dB
Harmonics < -10 dBc
Spurs < -80 dBc > 5 kHz offsets 500-1500 MHz