4-7
Using the HP/Agilent 83210A Service Kit
Making Measurements
3. Remove the correct coax jumper and connect a measurement instrument as shown in
thefollowing diagram. To measure signals going TO the module, measurements should
be made on the lower row of connectors on the extender module. Outputs coming
FROM the modules (going into the instrument) are measured on the top row of
connectors on the extender board.
4. Turn off the instrument’s power switch. Remove the module from the instrument.
Install the module onto the extender board and install the extender board into the
instrument.
5. Power on the instrument and make the measurements.