12-48
Module I/O Specifications
A23 Input (Agilent 8922M/S Only)
Outputs
Measure using known reference signal, refer to "A11 Receiver Mixer", page 12-19, for
procedure.
From Front Panel
AUX RF OUT J2
Freq Range 20 to 1000 MHz
Relative path loss with respect to siggen input,thru path (0 dB).
20 MHz Loss < 6 dB
1000 MHz Loss < 8 dB
Relativepath loss with respect to siggen input,atten 5 to 125 dB.
20 MHz Loss < 3 dB
1000 MHz Loss < 3 dB
To A19 Measurement Assembly
DET LO J6(14)
Meas Freq Range 90 to 1000 MHz
Output level (Uncal)
OFFSET VOUT (LOW) 100 mV +- 50 mV No input Power.
VOUT (LOW) 280 mV +- 50 mV + OFFSET (+10 dBm, 50 MHz)
VOUT (HIGH) .53 * V(LOW)
To A11 Receiver Mixer
1st_MIX_IN J5
Freq Range 20 to 1000 MHz
Output Level Normal 12 dBm to 22 dBm
Underrange 22 dBm to50 dBm