Localizing the Problem

If Power-Up Happened Correctly

If Power-Up Happened Correctly
If power-up happened correctly and no problem is indicated, this section is used to func-
tionally check most of the hardware. The generators are checked first with external mea-
surements, then the analyzers are checked with the generator. The RF Generator is
checked at 935 MHz and 10 dBm. The AF Generator is checked at 1 kHz and 1 V. These
checks are for indication only, performance tests in Chapter 3, “Verifying Performance”,
will test specifications.
NOTE If you possess an Agilent 8922S or Agilent 8922M, you should first re-configure your
instrument as an HP/Agilent 8922E or HP/8922G. To do this, select the following keys:
CONFIG (this is accessible from the Cell Control screen in the bottom right-hand
corner).
Compatible, select (HP 8922E or HP 8922G)
Referring to Figure 1-4, ensure the connections are made.
Figure 1-4 Front Panel Connections
Press .
NOTE On the HP/Agilent 8922A/B, press .
On the HP/Agilent 8922E/F/G/H/M/S, press , (RFG/RFA).
PRESET
AF To Oscilloscope
RF To Spectrum Analayzer
AUDIO OUT
AUDIO RF OUT
RF GEN/RF ANL
RF GEN/RF ANL
SHIFT CELL CNTL