1-5
Localizing the Problem
If Power-Up Checks FAILED
If Power-Up Checks FAILED
If the power up checks failed, continue with this section.
If the fan did not start, see "Troubleshooting the Power Supply", Chapter6.
Ifthe fan started, but any of the other power-up checks failed, see "Power-Up Self Test
Diagnostics".
If an error message occurs, refer to theAgilent 8922x Users Guide for additional
information.

Error Message Numbers

If the error message refers to a self test error it will be of the form:
One or more self tests failed: Error Code XXXX
Where xxxx corresponds to the error message number shown in the table below.
Table 1-1 Error Message Numbers
Failure Error
Number Suspect Assembly
Fatal Error - Host Processor Failure 0002 A7 Controller
Fatal Error - ROM Checksum Failure 0004 A8 Memory
Fatal Error - RAM Failure 0008 A8 Memory
Fatal Error - RAM Failure 0010 A8 Memory
Fatal Error - Timer Failure 0020 A7 Controller
Real Time Clock Failure 0040 A8 Memory
Keyboard Failure 0080 A1 Keyboard
Serial I/O Failure 0100 A21 GPIB
Internal Serial Bus Communication Failure 0200 Serial Bus
CRT Failure 0400 A19 CRT Drive
Miscellaneous Hardware Failure 0800 Miscellaneous H/W