12-33
Module I/O Specifications
A16 Receiver
Outputs
Figure 4-8 Typical Display
To test Pulse Demod apply RF Carrier with AM modulation to RF Input, measure Pulse
Demod Out on oscilloscope≅ 180 mV x %MOD
NOTE: Ensure correct settings on RF Analyzer page (frequency/amplitude). If the RF Input level
is greater than 5 dBm below RF Analyzer setting, the measurement will not register.
To A9 Global Test and Demod
UNMUTED_FM J1(14)
Sensitivity into 100 k load: 20µV/Hz See Figure 4-8on
page12-33
Max Deviation: 100 kHz peak
Amplitude 1.8 Vdc
To Front Panel
PULSE_DEMOD J1(7)
Level Pulse ON into open circuit: +2 V
Level Pulse OFF: 0 V
Output Impedance: 600
10-90% Rise/Fall time: <2.5µs