12-46
Module I/O Specifications
A23 Input (HP/Agilent 8922A.B,E,F,G,H) Only
To A19 Measurement Assembly
IN_VOLT J6(15)
AUTO_RNG_DET Prescaler AGC RF peak detector voltage.
Output Level 4.71± .5 V When AGC has active control.
AUTO_RNG_ALC Prescaler AGC modulator control voltage.
Range 0 to + 4 V
TEMP_DET Temperature sensor voltage.
Nominal Output 2.98± .1 Volts @ 25 Deg C
Sensivitity 10 mV / C
DUPLEX_DET Duplex port RF peak detector Voltage.
Nominal 100 mV± 20 mV @ +10 dBm
Trip Level 400 mV± 10 mV with relay closed.
785 mV± 10 mV with relay opened.
ANT_DET Antenna port RF peak detector voltage.
Nominal 100 mV± 20 mV @ +10 dBm
Trip Level 400 mV± 10 mV
FILTER_OUT_DET Receiver Output Port RF peak det. Voltage.
Nominal 10 mV± 5 mV @ -10 dBmOutput.
DET_HIGH RF Power peak detector high level voltage.
Output See DET HI/LO specs (RF Power Detector).
GND Input Section Analog Ground.
Nominal ± 10 mV