Diagnostics Theory

RF_DIAGS

RF Analyzer Loop 1 MHz Reference Detector

This test checks for the presence of the 1 MHz reference.

RF Analyzer Loop Lock Detector

This test checks the loop for lock at several frequencies.

Loop B Output Detector

This test checks the level detector at several frequencies.

Spectrum Analyzer

This test checks the A18 Spectrum Analyzer assembly.

Detector Output

The RF generator is routed externally to the spectrum analyzer through the AUX RF OUT and AUX RF IN front panel connectors. The level is set to a very low level (-100 dBm), and the spectrum analyzer detector is measured.

Stepped Gain

The stepped gain amplifiers are check using a 0 dB measurement as a reference. The steps are then measured one at a time using the detector.

IF Bandwidth

The IF bandwidth is set to all possible values and the voltage is measured by the detector.

Filter Rejection

The LO frequency is set at 10 times the bandwidth away from the center frequency of each filter, then the detector is read.

Variable Gain IF Amplifier

The variable IF gain amplifier is checked by turning on one bit of the control DAC at a time, starting with the LSB. The detector is checked as each bit is turned on.

RF Input Signal

With the internal calibration signal disabled, the RF generator and RF analyzer are set to 100 MHz for an IF of 114.3 MHz at the input of the spectrum analyzer. The signal is routed internally through the coupler on the RF IN/OUT port of the A23 Input assembly and the level is measured.

Receiver

This test checks the A16 Receiver assembly.

15-9

Page 223
Image 223
Agilent Technologies 8922 manual Spectrum Analyzer, Receiver

8922 specifications

Agilent Technologies 8922 is a versatile and robust mobile communication test set, designed to fulfill the demanding requirements of wireless device manufacturers and service providers. Primarily focused on mobile radio communications, the 8922 provides a comprehensive solution for testing a wide array of signaling standards and technologies, making it essential in the rapidly evolving telecommunications landscape.

One of the standout features of the Agilent 8922 is its support for multiple wireless standards, including GSM, GPRS, EDGE, WCDMA, and HSPA. This makes the device an ideal choice for testing mobile devices that operate on different generations of cellular networks. Additionally, the 8922 is equipped with advanced measurement capabilities, allowing users to perform key performance indicators assessments, such as power, modulation quality, and spectral emission analysis, ensuring devices meet stringent regulatory and performance standards.

The 8922 is designed with an intuitive user interface that simplifies the setup and execution of tests. It features a large color display, making it easier for users to view test results and navigate through various functions. The built-in automation capabilities enable seamless integration into production environments, where efficiency and repeatability are crucial for high-volume testing.

Technologically, the Agilent 8922 incorporates advanced signal processing algorithms, which enhance its measurement accuracy and speed. The device's architecture allows for real-time analysis of signals, enabling users to identify issues quickly and efficiently. This is particularly important in the realm of mobile communications, where time-to-market is critical for competitive advantage.

Another significant characteristic of the 8922 is its flexible configuration options. Users can customize the test set with different hardware and software modules to suit specific testing needs. This scalability ensures that the 8922 remains relevant as wireless technologies evolve and new standards emerge.

In conclusion, the Agilent Technologies 8922 is an indispensable tool for professionals engaged in the testing of mobile communications devices. Its broad support for multiple standards, advanced measurement capabilities, user-friendly design, and flexibility make it a top choice in the telecommunications testing community. As the mobile communication industry continues to advance, the 8922 stands out as a leading solution for ensuring that devices meet quality and performance expectations.