11-2

Self-Test Error Messages
Introduction
Introduction

Thischapter lists the error messages that appear on the status line of the display and on the

message screen when the self-tests are run on power-up. This list does not include all of

the messages that can appear under all circumstances.

”Battery Backed RAM Initialized. Structures corrupt.”
”Battery Backed RAM Initialized. Option RAM not maintained.”
”Battery Backed RAM Initialized.Optional RAM not found.”
”Battery Backed RAM Initialized. Standard RAM not maintained.”
”All host processor self-tests passed.”
”PANIC - UNKNOWN ERROR OCCURRED.”
”Attempt to write EEPROM failed.”
”Communication failure with the Rcvr Step Loop Board.”
A serial communications failure occurred with the A17 Step Loop assembly.
”Communication failure with the Sig Gen Step Loop Board.”
A serial communications failure occurred with the A26 Step Loop A assembly.
”Communication failure with the NSM/PMF Board.”
A serial communications failure occurred with the A5 Premod Filter and NSM assembly.
”Hop Controller communication channel Failure.”
”Self-test failure in Hop Controller.”
The A33 Hop Controller assembly failed its self-test.
”Hop Controller did not post self-test results.”
”DSP Analyzer communication Channel Failure.”
”Self-test failure in DSP Analyzer.”
The A9 Global Test and Demod assembly failed its self-test
”DSP Analyzer did not post self-test results.”
”Protocol Processor Communication Channel Failure.”
”Self-test failure in Protocol Processor.”
The A32 GSM Controller assembly failed its self-test.