276 Chapter 5
Language Reference
[:SENSe]:CORRection Subsection
[:SENSe]:CORRection Subsection

Delete All Corrections

[:SENSe]:CORRection:CSET:ALL:DELete
This command deletes all existing corrections.
History: Added with firmware revision A.08.00.
Front Panel
Access: Amplitude/Y Scale, Corrections, Delete All Corrections

Perform Amplitude Correction

[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Turns On or Off the amplitude corrections. When turned On, only the correction
sets that were turned on are enabled. When turned Off, all of the correct ion sets are
disabled.
Factory Preset
and *RST: Off
Remarks: To turn On or Off an individual correction set, use:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe].
Front Panel
Access: Amplitude/Y Scale, Corrections, Antenna, Correction On
Off
Amplitude/Y Scale, Corrections, Cable, Correction On Off
Amplitude/Y Scale, Corrections, Other, Correction On Off
Amplitude/Y Scale, Corrections, User, Correction On Off

Set Amplitude Correction Data

[:SENSe]:CORRection:CSET[1]|2|3|4:DATA
<freq>,<rel_ampl>{,<freq>,<rel_ampl>}
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Sets the amplitude correction data. These frequency/amplitude corrections will be
applied to the displayed data to correct for system losses/gains outside the
analyzer. Four different sets of correction data can be stored.
Example: :CORR:CSET1:DATA
900E6,0.3,1.0E9,0.35,1.3E9,0.2
Range: 200 points per set