Chapter 5 269
Language Reference
[:SENSe]:AVERage Subsection
[:SENSe]:AVERage Subsection

Clear the Current Average

[:SENSe]:AVERage:CLEar
Re-start the trace averaging fu nction.
NOTE Re-start the trace at the beginning of a sweep to obtain valid average data. To
do this, remotely abort the sweep and initiate a single sweep.

Set the Average Count

[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Specifies the number of measurements that are comb ined.
Factory Preset
and *RST: 100
Range: 1 to 8192
Front Panel
Access: BW/Avg, Average On Off

Turn Averaging On/Off

[:SENSe]:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AVERage[:STATe]?
This command toggles averaging off and on. Averaging combines the value of
successive measurements to av erage out measurement variations.
Factory Preset
and *RST: Off
Remarks: When a measurement under the front panel MEASURE key is
started, this command is turned off for video averaging
([:SENSe]:AVERage:TYPE VIDeo). If this command is
turned on for video averaging when any of the MEASURE key
measurements are in progress, that measurement will be
stopped.
Front Panel
Access: BW/Avg, Average On Off