Language Reference

[:SENSe]:FREQuency Subsection

Stop Frequency

[:SENSe]:FREQuency:STOP <freq>

[:SENSe]:FREQuency:STOP?

Set the stop frequency.

Factory Preset

 

and *RST:

1.06 GHz

Range:

EMC E7401A: –80 MHz1 to 1.58 GHz

 

EMC E7402A: –80 MHz1 to 3.10 GHz

 

EMC E7403A: –80 MHz1 to 6.78 GHz

 

EMC E7404A: –80 MHz1 to 13.3 GHz

 

EMC E7405A: –80 MHz1 to 27.0 GHz

Default Unit:

Hz

Front Panel

 

Access:

FREQUENCY/Channel, Stop Freq

Frequency Synthesis Mode

[:SENSe]:FREQuency:SYNThesis 123

[:SENSe]:FREQuency:SYNThesis?

This command switches between two phase noise optimization modes. Mode 2 optimizes the analyzer for close-in phase noise. Mode 3 optimizes the analyzer for tuning speed. Mode 1 is not recommended for new designs.

This command is available for the following models only:

E7402A, E7403A, E7404A, E7405A

Factory Preset

 

and *RST:

3

History:

Added with firmware revision A.08.00.

Front Panel

 

Access:

AUTO COUPLE, PhNoise Opt

1. 10 Hz minimum in log sweep mode.

Chapter 5

293

Page 293
Image 293
Agilent Technologies Model  A.08.xx Stop Frequency, Frequency Synthesis Mode, SENSeFREQuencySTOP freq SENSeFREQuencySTOP?

Model A.08.xx specifications

Agilent Technologies has long been a leader in the field of measurement and analytical instrumentation, and their Model A.08.xx is a prime example of this expertise. This advanced instrument is designed for a wide array of applications, spanning from life sciences to chemical analysis, offering unparalleled precision and reliability to meet the demands of laboratory environments.

One of the main features of the A.08.xx model is its advanced measurement capabilities. The instrument boasts a high-resolution detector that provides exceptional sensitivity, allowing researchers to detect even trace levels of analytes in complex samples. The enhanced signal-to-noise ratio is particularly beneficial for users working with low concentrations, ensuring accurate results without the need for laborious sample preparation.

The A.08.xx is equipped with state-of-the-art technologies that significantly enhance its performance. One such technology is its multi-wavelength detection system, which allows simultaneous analysis of multiple compounds within a single run. This not only boosts efficiency but also reduces the time required for method development and validation. Additionally, the model utilizes sophisticated software for data analysis, providing users with intuitive tools to interpret results quickly and effectively.

Another characteristic of the A.08.xx is its robust build quality and user-friendly interface. Designed for rigorous laboratory use, the instrument can withstand the demanding conditions of a busy research environment. Its intuitive touchscreen display simplifies operations, allowing users to set up experiments and navigate through various functions with ease. This user-centric design reduces the learning curve for new operators, enhancing productivity in the lab.

The A.08.xx also incorporates connectivity features that align with modern laboratory needs. With options for remote monitoring and data sharing, researchers can easily collaborate and access results in real-time, streamlining workflows and promoting innovation.

In conclusion, Agilent Technologies’ Model A.08.xx is not just an analytical instrument; it is a comprehensive solution for researchers and scientists seeking reliability and performance in their analytical work. With its advanced measurement capabilities, cutting-edge technologies, and user-friendly design, it continues to set the standard for excellence in laboratory instrumentation, facilitating groundbreaking research across various scientific fields.