Chapter 2 79
Status Registers
Use Status Registers to Determine the State of AnalyzerEvents and Conditions
Questionable Status Power Condition and Event Registers
The Questionable Status Power Condition Register continuously monitors output
power status of the analyzer. Condition registers are read-only. To query the
condition register, send the command
:STATus:QUEStionable:POWer:CONDition? The response will be the
decimal sum of the bits which are set to 1.
The transition filter specifies which types of bit state changes in the condition
register will set corresponding bits in the event register. The changes may be
positive (from 0 to 1) or negative (from 1 to 0). Send the comman d
:STATus:QUEStionable:POWer:NTRansition <num> (negative
transition) or :STATus:QUEStionable:POWer:PTRansition <num>
(positive transition) where <num> is the sum of the decimal values of the bits you
want to enable.
The Questionable Status Power Event Register latches transition events from the
condition register as specified by the transition filters. Event registers are
destructive read-only. Reading data from an event r egister will clear the content of
that register. To query the event register, send the command
:STATus:QUEStionable:POWer[:EVENt]?
See Questionable Status Event Enable Register on page77 for an explanation of
how to set the summary bit using the event enable register. In this case, use the
command :STATus:QUEStionable:POWer:ENABle <num>.
7 128 Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
8 256 CALibration Summary: This is the summary bit fo r the
Questionable Calibration Status Register.
9 512 INTegrity Sum: This is the summary bit for the
Questionable Integrity Status Register.
10 1024 Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
11 2048 Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
12 4096 Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
13 8192 Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
14 16384 Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
15 32768 Always Zero (0)
Bit Decimal
Valu e
Description