Language Reference

CALCulate:EMI:SLISt Subsection

 

 

The ADD command always appends signals to the end of the signal list. If the

Table 5-1

Field Description Table

 

 

 

 

 

 

 

 

 

Position

ADD

ADD

Definition

Default

 

 

Required

Optional

 

 

 

 

 

 

 

 

 

1

Yes

 

Frequency

Value required

 

 

 

 

 

 

 

2

 

Yes

Peak Ampl

0

(Peak detector flag off)

 

 

 

 

 

 

 

3

 

Yes

QPK Ampl

0

(Quasi-Peak detector flag off)

 

 

 

 

 

 

 

4

 

Yes

Avg Ampl

0

(Average amplitude flag off)

 

 

 

 

 

 

 

5

 

Yes

Uncertainty

0

(Applied when measured)

 

 

 

 

 

 

 

6

 

Yes

Total

0

(Applied when measured)

 

 

 

 

Correction

 

 

 

 

 

 

 

 

 

7

 

Yes

Marked Status

0

(off)

 

 

 

 

 

 

 

8

 

Yes

Peak Det. Flag

0

(off) * Implicitly On if peak

 

 

 

 

 

amplitude provided

 

 

 

 

 

 

 

9

 

Yes

QPK Det. Flag

0

(off) * Implicitly On if QPk

 

 

 

 

 

amplitude provided

 

 

 

 

 

 

10

 

Yes

Avg Flag

0 (off) * Implicitly On if Avg amplitude

 

 

 

 

 

provided

 

 

 

 

 

 

11

 

Yes

Comment

Empty

 

 

 

 

 

 

 

ADD command fails (e.g. signal list at capacity), a message is placed on the status line and in the error queue. An invalid data format places a message on the status line and into the error queue.

Because all of the values may not be meaningful during the ADD process, the ADD command allows defaults to be used. Because of this the data string submitted may be abbreviated. Each ADD data format string must contain a valid frequency string. The rule applied to ADD strings is that values need to be provided only as far to the right as a non-default value is used.

For example, the data format to add a signal at 1 MHz can be indicated as a frequency only. The two commands below are equivalent. The remaining fields are set to their default values.

CALCulate:EMI:SLISt:ADD “1e6” or

CALCulate:EMI:SLISt:ADD “1000000”

If the frequency and a comment are desired, the missing fields between the frequency and comment position must be shown by marking the positions using the comma separator. Either type of quote delimiting is available. The preceding example would be expressed as:

CALCulate:EMI:SLISt:ADD “1e6,,,,,,,,,,This is my comment” or

CALCulate:EMI:SLISt:ADD '1e6,,,,,,,,,,This is my comment’

202

Chapter 5

Page 202
Image 202
Agilent Technologies Model  A.08.xx manual Field Description Table, CALCulateEMISLIStADD 1e6 or

Model A.08.xx specifications

Agilent Technologies has long been a leader in the field of measurement and analytical instrumentation, and their Model A.08.xx is a prime example of this expertise. This advanced instrument is designed for a wide array of applications, spanning from life sciences to chemical analysis, offering unparalleled precision and reliability to meet the demands of laboratory environments.

One of the main features of the A.08.xx model is its advanced measurement capabilities. The instrument boasts a high-resolution detector that provides exceptional sensitivity, allowing researchers to detect even trace levels of analytes in complex samples. The enhanced signal-to-noise ratio is particularly beneficial for users working with low concentrations, ensuring accurate results without the need for laborious sample preparation.

The A.08.xx is equipped with state-of-the-art technologies that significantly enhance its performance. One such technology is its multi-wavelength detection system, which allows simultaneous analysis of multiple compounds within a single run. This not only boosts efficiency but also reduces the time required for method development and validation. Additionally, the model utilizes sophisticated software for data analysis, providing users with intuitive tools to interpret results quickly and effectively.

Another characteristic of the A.08.xx is its robust build quality and user-friendly interface. Designed for rigorous laboratory use, the instrument can withstand the demanding conditions of a busy research environment. Its intuitive touchscreen display simplifies operations, allowing users to set up experiments and navigate through various functions with ease. This user-centric design reduces the learning curve for new operators, enhancing productivity in the lab.

The A.08.xx also incorporates connectivity features that align with modern laboratory needs. With options for remote monitoring and data sharing, researchers can easily collaborate and access results in real-time, streamlining workflows and promoting innovation.

In conclusion, Agilent Technologies’ Model A.08.xx is not just an analytical instrument; it is a comprehensive solution for researchers and scientists seeking reliability and performance in their analytical work. With its advanced measurement capabilities, cutting-edge technologies, and user-friendly design, it continues to set the standard for excellence in laboratory instrumentation, facilitating groundbreaking research across various scientific fields.