Agilent Technologies Model A.08.xx manual Data Uncalibrated Summary This is the summary

Models: Model A.08.xx

1 349
Download 349 pages 21.69 Kb
Page 90
Image 90

Status Registers

Use Status Registers to Determine the State of Analyzer Events and Conditions

Bit

Decimal

Description

 

Value

 

 

 

 

0

1

Reserved: This bit is not used by the analyzer, but is for

 

 

future use with other Agilent products.

 

 

 

1

2

Reserved: This bit is not used by the analyzer, but is for

 

 

future use with other Agilent products.

 

 

 

2

4

Reserved: This bit is not used by the analyzer, but is for

 

 

future use with other Agilent products.

 

 

 

3

8

Data Uncalibrated Summary: This is the summary

 

 

bit for the Questionable Status Integrity Uncalibrated

 

 

Register.

 

 

 

4

16

IF/ADC Over Range: The signal input level is too high,

 

 

causing the analyzer analog-to-digital converter (ADC)

 

 

range to be exceeded. This may occur with resolution

 

 

bandwidths less than or equal to 300 Hz in zero span.

 

 

 

5

32

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

6

64

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

7

128

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

8

256

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

9

512

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

10

1024

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

11

2048

Reserved: This bit is not used by the analyzer, but are for

 

 

future use with other Agilent products.

 

 

 

12

4096

Invalid Data: A 1 in this bit position indicates that the

 

 

present trace data does not reflect the existing analyzer

 

 

state. Trigger a new sweep and/or measurement.

 

 

 

13

8192

Reserved: This bit is not used by the analyzer, but is for

 

 

future use with other Agilent products.

 

 

 

14

16384

Reserved: This bit is not used by the analyzer, but is for

 

 

future use with other Agilent products.

 

 

 

15

32768

Always Zero (0): This bit is always set to 0.

 

 

 

90

Chapter 2

Page 90
Image 90
Agilent Technologies Model  A.08.xx manual Data Uncalibrated Summary This is the summary

Model A.08.xx specifications

Agilent Technologies has long been a leader in the field of measurement and analytical instrumentation, and their Model A.08.xx is a prime example of this expertise. This advanced instrument is designed for a wide array of applications, spanning from life sciences to chemical analysis, offering unparalleled precision and reliability to meet the demands of laboratory environments.

One of the main features of the A.08.xx model is its advanced measurement capabilities. The instrument boasts a high-resolution detector that provides exceptional sensitivity, allowing researchers to detect even trace levels of analytes in complex samples. The enhanced signal-to-noise ratio is particularly beneficial for users working with low concentrations, ensuring accurate results without the need for laborious sample preparation.

The A.08.xx is equipped with state-of-the-art technologies that significantly enhance its performance. One such technology is its multi-wavelength detection system, which allows simultaneous analysis of multiple compounds within a single run. This not only boosts efficiency but also reduces the time required for method development and validation. Additionally, the model utilizes sophisticated software for data analysis, providing users with intuitive tools to interpret results quickly and effectively.

Another characteristic of the A.08.xx is its robust build quality and user-friendly interface. Designed for rigorous laboratory use, the instrument can withstand the demanding conditions of a busy research environment. Its intuitive touchscreen display simplifies operations, allowing users to set up experiments and navigate through various functions with ease. This user-centric design reduces the learning curve for new operators, enhancing productivity in the lab.

The A.08.xx also incorporates connectivity features that align with modern laboratory needs. With options for remote monitoring and data sharing, researchers can easily collaborate and access results in real-time, streamlining workflows and promoting innovation.

In conclusion, Agilent Technologies’ Model A.08.xx is not just an analytical instrument; it is a comprehensive solution for researchers and scientists seeking reliability and performance in their analytical work. With its advanced measurement capabilities, cutting-edge technologies, and user-friendly design, it continues to set the standard for excellence in laboratory instrumentation, facilitating groundbreaking research across various scientific fields.