270 Chapter 5
Language Reference
[:SENSe]:AVERage Subsection
Turn Automatic Averaging On/Off
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
Sets the averaging to be automatically set to the appropriate type for the current
measurement setup. Or allows you to manually choose the type of averaging with
[:SENSe]:AVERage:TYPE.
When AUTO is On:
If the Y Axis Scale is not L inear or Log, then average type i s Video (Y Axis
Scale) Averaging.
If the Y Axis Scale is Linear or Log, then average type is Pow er Averaging.
If the Detector is Peak, Sample, or Negative Peak (not Average), then aver age
type is Video Average.
See Figure 5-2, which shows these auto rules for average type in flowchart format.
Figure 5-2 Auto Rules for Average Type