Respond to alarm message: (con’t)
Alarm Message Urgency What It Means What to Do
High – H, Medium – M, Intermediate – I, Low – L
Cal Required: Redo BG
in 2 hrs.
ISensor signal is settling. e system requires another
calibration in 2 hours. You will continue to get
glucose results during that time.
Wait for the “Do BG test” prompt and perform
another Blood Glucose mode test for calibration.
You will have half hour to perform the calibration
aer which you will not get glucose results
or alarms until the calibration is completed
successfully.
Cool Down Skin. LUnable to calibrate the system because the skin
temperature is too warm.
Cool the skin at the insertion site and wait for the
“Do BG Test” prompt and perform another Blood
Glucose mode test for calibration. You can cool
down your skin temperature by moving to a cooler
environment. If you are wearing warm clothing
(like a sweater or jacket), try taking them o.
Cool Down Skin Redo
BG Later.
ICalibration failed because your skin temperature is
too warm.
Cool the skin at the insertion site. Wait for half an
hour and perform another Blood Glucose mode
test for calibration. You can cool down your skin
temperature by moving to a cooler environment. If
you are wearing warm clothing (like a sweater or
jacket), try taking them o.
Did You Remove e
Sensor?
L•esystemhasdetectedthatthesensorwasjust
removed.
•Ifthismessageappearsandyoudidnotremove
the sensor, there may be a faulty connection
between the sensor and the transmitter.
•Ifyoudidnotjustremovethesensor,selectNo.
•Ensurethatthetransmitterisattachedrmlyto
the sensor support mount.
•Ensurethatthesensorsupportmountisattached
rmly to the skin.
•Ifthealarmpersists,removethesensorandinsert
a new one (see Section 5).
Customer Care: 1-866-597-5520
94 Section 10
Respond to Alarms, Errors, and Problems