8
OVER
8 Set, Review, or Change the Alarm Settings
Purpose To set the receiver to:
• Alertyouwhenyourglucoseconditionschange.
• Notifyyouwhenthereceiverneedsattention.
• Notifyyouwhenthereceiverbeginsorcompletescertainactivities.
Start Do this task:
• ImmediatelyaeryounishtherstsetupofyourFreeStyleNavigatorsystem.
• Whenyourhealthcareteamgivesyounewglucosethresholdnumbers.
• Anytimeyougeta“PreferencesLostRecongure”alarmmessageonthereceiver.
• Anytimeyoureplaceyourreceiver.
• Anytimeyouwanttochangeyouralarmsettings.
• Anytimeyouwanttoreviewyouralarmsettings.
Materials • YourFreeStyleNavigatorReceiver.
 • Lowandhighglucosethresholdvalues(fromyourhealthcareteam).
Helpful Information
•High and low glucose alarms let you know when your glucose levels are above or below your low and
high glucose threshold values. You will decide the low and high glucose threshold values based on
input from your healthcare team. If there is no glucose data, you will not get glucose alarms.
•Aprojected alarm is an alarm that provides an early warning of an event that is likely to occur if the
current trend continues. Projected alarms use the low and high glucose threshold values regardless of
whether or not the low and high glucose alarms are turned on.
For example, a projected low alarm lets you know when you are approaching your low glucose threshold
value. You will decide the amount of notice required (whether you should be notied approximately 10
minutes, 20 minutes or 30 minutes in advance of reaching your threshold value) based on the input from
your healthcare team.
•Adatalossalarmwarnsyouwhenglucoseresultsarenolongeravailable.eyoccurwhenthesensor
has expired, when the transmitter-receiver connection is broken, when a calibration has expired or
when the sensor is not working properly. e data loss alarm cannot be turned o without turning o
the glucose alarms (Low Glucose, High Glucose, Projected Low Glucose and Projected High Glucose).
•System alarms notify you of events like low battery or time to calibrate.
•Progress tones notify you of errors, results, and successful completion of steps in activities like Blood
Glucose mode testing.
Caution:
e high and low alarms are intended to assist you in managing your diabetes and should
not be exclusively used to detect hypoglycemia or hyperglycemia. e alarms should always be used in
conjunction with other indications of glycemic state such as your glucose level, trend, line graph etc.
Caution:
High and low glucose alarms are DIFFERENT from your glucose targets. Low and high glucose
alarms alert you when you’ve crossed a certain low or high value. Glucose targets allow the reports and
line graphs to show how your glucose levels have been performing compared to your set targets.
Caution: Before adjusting treatment for diabetes management based on the continuous glucose results
from your FreeStyle Navigator system, perform a Blood Glucose mode test to conrm the continuous
result.

Section 8 67

Set, Review, or Change the Alarm Settings