Respond to alarm message: (con’t)
Alarm Message Urgency What It Means What to Do
High – H, Medium – M, Intermediate – I, Low – L
Transmitter Error
Alarms Inactive.
ITransmitter problem.
Alarms are not working.
Wait for the problem to go away. If the problem
persists aer 50 minutes, you will get the alarm.
“Alarms Inactive. Replace Transmitter”.
TX Battery Exhausted
Replace Battery.
LTransmitter battery power loss is imminent. Remove sensor, replace transmitter battery,
reconnect transmitter and receiver, and insert a
new sensor.
Warm Up Skin. LUnable to calibrate the System because the skin
temperature is too cold.
Warm the skin at the insertion site and wait for the
“Do BG Test” prompt and perform another Blood
Glucose mode test for calibration. You can warm
up the skin temperature by moving to a warmer
environment or by putting on some warm clothing.
Warm Up Skin Redo
BG Later.
ICalibration failed because your skin temperature is
too cold.
Warm the skin at the insertion site. Wait for the
blood drop icon to appear in about half an hour
and perform another Blood Glucose mode test for
calibration. You can warm up the skin temperature
by moving to a warmer environment or by putting
on some warm clothing.
Section 10 99 Respond to Alarms, Errors, and Problems