7
OVER
7 Calibrate Your System/Test Blood Glucose Manually
Purpose To ensure the receiver will calculate your glucose readings accurately.
Start Do this procedure:
•Eachtime the receiver prompts you with the message, “Do BG Test.” You will
see a blood drop icon on the screen and you will also hear the system beep
(or vibrate).
•Whenyouneedtomeasureyourbloodglucosemanuallyusingthebuilt-in
FreeStyle Meter.
Materials •Soap,water,andacleantowel. •FreeStyleTestStrips.
•AFreeStyleLancingDevice. •YourFreeStyleNavigatorReceiver.
•Sterilelancets. •Asharpscontainer.
Helpful Information
Important ings to Remember About System Calibration
e system is cali• brated using a capillary whole blood glucose measurement (performed in the Blood
Glucose mode using the built-in FreeStyle Blood Glucose Meter). In order to calibrate, the receiver
compares the reading it gets from the blood to the reading the sensor takes from the interstitial uid.
Based on the calibration, the receiver calculates the glucose readings.
Do not use venous blood for calibrating the system.•
Important: Your blood glucose level must be between 60 and 300 mg/dL (3.3 and 16.7 mmol/L) to be
able to perform calibration tests. If your blood glucose level is changing rapidly, you may not be able
to calibrate the system. For example, during a meal or exercise, your glucose levels may vary rapidly.
Try to time your sensor insertion so that your calibration times do not coincide with your regular
meal or exercise activities.
Caution:
Always calibrate the system only using a nger-stick blood sample. Do NOT use alternate
site blood glucose measurements to calibrate the system. e receiver contains a built-in FreeStyle®
Blood Glucose Meter for performing calibration tests.
Important: In clinical trials, we observed that the sensor signal sometimes temporarily decreases
from the true value. is typically happens at night during sleep, and recovers rapidly when the user
moves or is awakened. However, in order to avoid being aected by this phenomenon the system
should not be calibrated when the wearer is asleep.
You may not be able to calibrate the system if your glucose levels are changing rapidly (e.g. during •
or aer exercise, meals or insulin dosing). Under such conditions, the system may not ask you to
calibrate; instead, it will delay its request until conditions are acceptable.

Section 7 53

Calibrate Your System/Test Blood Glucose Manually