19 Appendix C: FCC Compliance Information (con’t)

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

Table 1. e FreeStyle Navigator system is intended for use in the electromagnetic environment specied below. e customer or the user of the

FreeStyle Navigator system should assure that it is used in such an environment.

Emissions Test Compliance Electromagnetic Environment – Guidance

RF Emissions
CISPR 11
Group 1 e FreeStyle Navigator system uses RF Energy only for its Intended function. erefore, its
RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B e FreeStyle Navigator system is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not Applicable

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

Table 2. e FreeStyle Navigator system is intended for use in the electromagnetic environment specied below. e customer or user of the FreeStyle

Navigator system should assure that it is used in such an environment.

Immunity test Test Level FreeStyle Navigator System

Compliance Level

Electromagnetic Environment– Guidance

Tx/Rx Electrostatic Discharge
(ESD)
IEC60601-1-2
±8 kV Contact Discharge
±15 kV Air Discharge
±8 kV Contact Discharge
±15 kV Air Discharge
Floors should be wood, concrete or ceramic tile. If oors
are covered with synthetic material, the relative humidity
should be at least 30%.
Tx/Rx Electrostatic Discharge
(ESD)
EN 300 489-3
±4 kV Contact Discharge
±8 kV Air Discharge
±4 kV Contact Discharge
±8 kV Air Discharge
Floors should be wood, concrete or ceramic tile. If oors
are covered with synthetic material, the relative humidity
should be at least 30%.
Magnetic Field Immunity
IEC 60601-1-2
3A/m @ 50Hz and 60Hz
5 min dwell time
3A/m @ 50Hz and 60Hz
5 min dwell time
Power frequency magnetic elds should be at a level
characteristic of a typical location in a typical commercial or
hospital environment.
Transmitter Radiated Emissions
FCC
CFR 47 Part 15.231(e)
Must be below
72.9 dBuV/m @433.56MHz
67.9 dBuV/m @ 433.56MHz
Customer Care: 1-866-597-5520
166 Section 19
Appendix C: Classification/Compliance/Declaration