OVER
Respond to alarm message: (con’t)
Alarm Message Urgency What It Means What to Do
Caution: Before adjusting treatment for diabetes management based on the continuous glucose results from your
FreeStyle Navigator system, perform a Blood Glucose mode test to conrm the continuous result.
High – H, Medium – M, Intermediate – I, Low – L
Projected Low Glucose. Mis is an early warning that your glucose level and
the rate at which it is changing indicate that you
may soon reach your low glucose threshold.
When set to High sensitivity, you have
approximately 30 minutes before reaching your low
glucose threshold.
When set to Medium sensitivity, you have
approximately 20 minutes before reaching your low
glucose threshold.
When set to Low sensitivity, you have
approximately 10 minutes before reaching your low
glucose threshold.
Follow the recommendations of your healthcare
team for treating or stopping the drop in glucose
that you are experiencing.†
Receiver Error Alarms
Inactive.
IReceiver error. e problem could go away in 50 minutes or less.
If problem persists, call Customer Care.
Reconnect. Le receiver has not been receiving signals sent by
the transmitter for 5 minutes. e receiver is either
too far from the transmitter or there are materials
or signals causing interference.
Be sure the receiver and transmitter are within
10 feet of one another. If the alarm persists,
select System from the Main menu, then select
Reconnect. See Section 9, Step 2.
Reconnect–Alarms
Inactive.
Me receiver has not been receiving signals sent by
the transmitter for 8 minutes. e receiver is either
too far from the transmitter or there are materials
or signals causing interference.
Be sure the receiver and transmitter are within
10 feet of one another. If the alarm persists,
select System from the Main menu, then select
Reconnect. See Section 9, Step 2.
Remove Any Sensor
Alarms Inactive.
IAlarms are not operating. e receiver thought the
sensor was inserted when the receiver was reset.
Remove the sensor and insert a new one (see
Section 5). Ignore the message if there is no
sensor inserted.
Section 10 97 Respond to Alarms, Errors, and Problems