B&K 4071 Int FM: 1,000 Hz, Pk dev: 10,000 Hz, Internal FM Mode Parameters, Level, 10.0dBm

Models: 4071

1 88
Download 88 pages 40.73 Kb
Page 21
Image 21
5.4 Internal FM Mode

5.4 Internal FM Mode

Introduction

Internal FM mode generates frequency-modulated signal of fixed amplitude. It used internally generated signal to modulate frequency of carrier signal. Modulated and modulating frequency, output level, deviation, and DC offset values could be manually entered from the front panel keypad in this mode. External signal gating options is available in this mode. Internal FM mode could be entered by pressing “Mode”> “FM” >”1” key at any time.

Internal FM Mode Parameters

The Internal FM mode has the following front panel display:

 

 

 

Int FM: 1,000 Hz

Pk dev: 10,000 Hz

 

 

1

 

 

 

 

 

 

 

3

 

 

1,000,000.00 Hz

-10.0 dBm

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 5.4-1: Internal FM mode display

1. Modulating Frequency

2

4

In this field enter the frequency of the modulating sinusoid. You may enter from 0 Hz to 10,000 Hz in 1 Hz steps.

2. Peak Frequency Deviation

In this field specify the degree to which the modulating signal is allowed to change the carrier frequency. You may enter from 0 Hz (no change) to 5.0 MHz in 1 Hz steps

This parameter is a peak value. If the deviation were 1 KHz and the carrier frequency were 1 MHz, for example, then the output frequency will swing between a maximum of 1 MHz + 1 KHz and a minimum of 1 MHz - 1 KHz.

NOTE: If values for the deviation and carrier frequencies are entered such that the output frequency exceeds the 0 to 31.5 MHz range, distortion of the output waveform may result.

3. Carrier Frequency

In this field enter the frequency of the carrier. You may enter from 0 Hz (DC) to 31.5 MHz in 0.01 Hz steps.

NOTE: If values for the deviation and carrier frequencies are entered such that the output frequency exceeds the 0 to 31.5 MHz range, distortion of the output waveform may result.

4. Level

In this field enter the output level, from 4 mVp-p to 5 Vp -p in 1 mV steps or from -44.0 dBm to +18.0 dBm in 0.1 dBm steps.

NOTE: The level specified is a 50O loaded level. This is the level of the signal, which will appear across a 50O load connected to the SIG Out connector. Into an open circuit, the output swing will be twice the value entered.

Offset

You can enter an offset voltage for the output signal. For more information on output offsets refer to section 4.4.

Ext Gating Input

This TTL compatible input can be used to gate the output signal on or off. A logic high voltage turns off the output. For further information on the Ext Gating Input, refer to section 2.0.

16

BK Precision 4071 User Manual Rev.2.2

Page 21
Image 21
B&K 4071 Int FM: 1,000 Hz, Pk dev: 10,000 Hz, Internal FM Mode Parameters, Peak Frequency Deviation, Level, 10.0dBm