System Configuration Manual
300510EN, Rev AA
December 2010
Rosemount Raptor
5-84 Section 5. Installing a Raptor Level Gauging System
5.9.2 Installing a 5400 Using the Installation
Wizard
The installation wizard in TankMaster WinSetup is a tool that may be used to
install and configure a Rosemount 5400 Radar Level Transmitter and other
devices.
However, in most cases the following procedure is recommended when
installing a new Rosemount 5400 transmitter:
1. Configure the Rosemount 2160 FCU Slave Database by including the
new 5400 transmitter. Ensure that the correct Modbus address is
configured (see “Installing a Rosemount 2160 FCU” on page 5-25 for
more information).
2. Configure the Rosemount 2410 Tank Database in the 2410 Tank Hub
Properties/Tank Database window, (see “Installing a Rosemount 2410
Tank Hub” on page5-38 for more information).
3. Install the 5400 in TankMaster via the 2410 Tank Hub Properties/Device
Tags window.
4. Configure the 5400 (see “Configuration via 5400 Properties” on
page 5-79).
See also “Adding a Tank to a Raptor System” on page5-114 for further
information on adding tanks and devices to a Raptor system.
In case the recommended installation procedure as described above can not
be used, the installation wizard offers an alternative method to install a new
Rosemount 5400 Radar Level Transmitter in TankMaster. In this case do the
following:
1. Configure the Rosemount 2160 FCU Slave Database by including the
new 5400 transmitter. Ensure that the right Modbus address is
configured (see “Installing a Rosemount 2160 FCU” on page 5-25 for
more information).
2. Configure the Rosemount 2410 Tank Database in the
2410 TankHub Properties/Tank Database window, (see “Installing a
Rosemount 2410 Tank Hub” on page5-38 for more information).
3. Install and configure the Rosemount 5400 as described in “Using the
installation wizard” on page 5-85.