ICE CONTROL SOFTWARE ICS62XX

2.3.10Measurement during command execution

The ICS62XX possesses a counting function which counts the time or the number of steps from starting the target program to the occurrence of a break.

The counting range is described below.

(1) Time counting mode

6.5 µsec to 6.5 ⋅ 65535 µsec (=425.977 msec)

Measurement error : ±6.5 µsec

(The display is in millisecond units: msec)

(2)Step counting mode

Step 1 to step 65535

Measurement error : 0 steps

(error of 1 step may be presumed during interrupt process)

When the measurement range is exceeded, the following message is displayed:

*RUN TIME=TIMEOVER.

2.3.11Self-diagnostic function

The ICE performs a self-check at power ON. When a check instruction (#CHK) is input from the host system, the self-test results are sent to the host.

#CHK

#...System awaits instruction unless an error occurs.

A check instruction is automatically input when the ICS62XX system program is loaded.

B>ICS62XX

* ICE POWER ON RESET * * DIAGNOSTIC TEST OK *

#

(Epson logo appears)

(Check instruction is automatically input; if no anomaly occurs, the following message appears)

When the above display appears, it indicates that the ICE and host are connected properly and the ICE is operating correctly.

If the ICE is power supply is OFF or the the cable to the host is not connected at the prompt, the following message appears:

B>ICS62XX

*COMMUNICATION ERROR OR ICE NOT READY*

Then, when the ICE power supply is switched ON, a self-test is automatically performed and the following message is displayed:

* ICE POWER ON RESET * * DIAGNOSTIC TEST OK *

#

When an error message is displayed after entering the check instruction, it is likely to be due to hardware failure. Contact customer support.

VIII-16