Advanced Operations

Chemical form:

Enter the chemical formula used in the display to represent the analysis parameter.

Use the alphanumeric keypad to enter the chemical form and press Next to enter Calibration settings.

6.1.1.3 Calibration settings for single and multi wavelength mode

 

A method is calibrated by determining the absorbance values of

 

several standard solutions of known concentration.

 

There are three ways to create and store a calibration curve.

 

Instructions for each method follows (Table 9).

 

Table 9 Calibration settings

 

 

Mode

Descriptions

 

 

 

A calibration table is created by entering the concentration values and the absorbance values of

Enter values

the analyte solution. The absorbance values are plotted versus standard concentrations and the

 

calibration curve is displayed as a graph (page 66).

 

A calibration table is created by entering the concentration values of the standard solutions and

Read Standards

then measuring the absorbance of the analyte solutions. The absorbance values are plotted

 

versus standard concentrations and the calibration curve is displayed as a graph (page 68).

 

 

 

If the calibration curve can be determined from the mathematical relationship between

Enter Formula

concentration and absorbance by linear regression, etc., the corresponding formula can be

selected (linear, 2nd or 3rd order polynomial) from a list and the appropriate factors can be

 

 

entered (page 69).

 

 

 

Calibration by entering calibration values

 

Enter concentration/absorbance

 

1. Select the Enter Values and press Next.

66

Page 66
Image 66
Hach DR 2800 Calibration settings for single and multi wavelength mode, Calibration by entering calibration values