Standard Operations

1.Press Options>More>Chemical Form.

2.Select the Chemical Form.

Note: The stoichiometric conversion of the measurement result is carried out automatically.

Note: The selected Chemical Form will appear on the display. Test results will be calculated and displayed in this chemical form.

5.4.6.1Change of the default setting of the chemical form

1.Insert the sample cuvette or blank (depending on the working procedure) into the cell compartment.

2.In the result display, press Options>More>Chemical Form.

3.A list of available evaluation forms appears. Select the new default setting.

4.Press Save as Default.

The current result and all further measurements will be displayed in the new chemical form.

5.4.7 Run a reagent blank

Some of the stored tests/methods include the "Reagent Blank" function. This enables the reagent blank value to be measured and then taken into account in calculating the measurement result.

Measurement/analysis of a reagent blank:

1.Prepare the test/method in accordance with the (working) procedure. Instead of a sample, deionized water is used to determine the reagent blank value.

2.Select the test. If required by the (working) procedure, insert the cell with deionized water into the cell compartment. Press Zero.

3.Insert the prepared sample cell into the cell compartment. Press Read. The result is displayed.