Section 1

Specifications

 

 

Specifications are subject to change without notice.

 

 

Performance Specifications

 

 

 

 

Operating Mode

 

Transmittance (%), Absorbance and Concentration

 

 

 

Source Lamp

 

Gas-filled Tungsten (visible)

 

 

 

Wavelength Range

 

340–900 nm

 

 

 

Wavelength Accuracy

 

± 1.5 nm

 

 

Wavelength Reproducibility

< 0.1 nm

 

 

 

Wavelength Resolution

 

1 nm

 

 

 

Wavelength Calibration

 

Automatic

 

 

 

Wavelength Selection

 

Automatic, based on method selection

 

 

 

Spectral Bandwidth

 

< 8 nm

 

 

Photometric measuring range

± 3.0 Ext in Wavelength Range 340–900 nm

 

 

 

Photometric Accuracy

 

5 mAbs at 0.0–0.5 Abs

 

1% at 0.50–2.0 Abs

 

 

 

 

 

 

 

< 0.5%–2 Abs

Photometric Linearity

 

< = 1% at > 2 Abs

 

 

with neutral glass at 546 nm

 

 

 

Stray Light

 

< 0.2% T @ 340 nm with KV450/3

 

< 0.1% T @ 340 nm with NaNO2

 

 

Data storage

 

500 measured values (result, date, time, sample ID, user ID)

 

 

 

User programs

 

50

 

 

Physical and Environmental Specifications

 

 

 

Width

 

220 mm (8.6 in)

 

 

 

Height

 

135 mm (5.3 in)

 

 

 

Depth

 

330 mm (12.9 in)

 

 

 

Weight

 

4.06 kg (8.95 Ibs) without battery

 

4.38 kg (9.66 Ibs) with battery

 

 

 

 

Operating Requirements

10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)

 

 

 

Storage Requirements

 

-40–60 °C (-40–140 °F) max. 80% relative humidity (non-condensing)

 

 

Additional technical data

 

 

 

 

Mains connection

 

15 V- / 30VA

 

Plug-in power supply unit: (100–240 V/50–60 Hz)

 

 

 

 

 

 

 

Use only shielded cable with max. length of 3 m.

Interfaces

 

1 x USB type A

 

 

1 x USB type B

 

 

 

Enclosure Rating

 

IP 41 with closed lid

 

IP 42 with Protective Cover in place

 

 

 

 

 

Safety class

 

Safety class II

 

 

 

7

Page 7
Image 7