6.7.3.3 Absorbance check

The Absorbance Check test is used to test the photometric accuracy and repeatability of the instrument.

This test can be used to test absorbance at any wavelength by a specific sample or test filter set (Refer to section 6.7.3.4 on page 102).

1.Press Absorbance Check in the Optical Checks menu.

2.Press λ to input the wavelength.

3.Input the wavelength and press OK to confirm.

4.Remove any cuvette/sample cell from the cell compartment and press Zero.

5.Insert sample cuvette/cell into the cell compartment and press Read.

6.5 Replicates of blanking and reading lead to results.

7.The result is displayed and is to be compared with the nominal/standard data of the sample cuvette/cell.

8.Press Cancel to return to Optical Checks.