Hach DR 2800 user manual Absorbance check, 101

Models: DR 2800

1 126
Download 126 pages 31.47 Kb
Page 101
Image 101

Advanced Operations

6.7.3.3 Absorbance check

The Absorbance Check test is used to test the photometric accuracy and repeatability of the instrument.

This test can be used to test absorbance at any wavelength by a specific sample or test filter set (Refer to section 6.7.3.4 on page 102).

1.Press Absorbance Check in the Optical Checks menu.

2.Press λ to input the wavelength.

3.Input the wavelength and press OK to confirm.

4.Remove any cuvette/sample cell from the cell compartment and press Zero.

5.Insert sample cuvette/cell into the cell compartment and press Read.

6.5 Replicates of blanking and reading lead to results.

7.The result is displayed and is to be compared with the nominal/standard data of the sample cuvette/cell.

8.Press Cancel to return to Optical Checks.

101

Page 101
Image 101
Hach DR 2800 user manual Absorbance check, 101