Advanced Operations
Peak volume/sample volume methods
1.Select Stored Programs in the main menu. Select the required program.
2.Press Start.
3.Analyze a sample without added standard solution in accordance with the instructions in the Procedures Manual. When the measurement is complete, leave the sample cuvette/cell in the cuvette/sample cell holder.
4.Press Options>More>Standard Addition.
An overview of the data of the standard addition procedure is displayed.
5.Press OK to accept the standard values for standard concentration, sample volume (total) and standard addition volume. Press Edit to change any of these values.
84