Advanced Operations
6.1.2 Free programming program type
Free Programming is an advanced option for entering original
Important Note: Perform steps
Method (section 6.1.1 on page 61) before proceeding with Free
Programming.
Highlight the line containing the program point that is to be edited or defined and press Edit.
| Table 10 Definitions of the program points |
|
|
Program point | Description |
|
|
Name | Name of the analysis parameter |
|
|
Version | An abbreviation or version number assigned by the user is entered here. |
|
|
| Exact definition of the test: the number of wavelengths at which measurements are made, the |
Measurement Process | number of absorbance measurements needed, the keys to be used, any waiting periods between |
| measurements, etc. |
Formula | Definition of the formulas with which the test result is calculated. |
|
|
| The number of variables shown in the display depends on the definition of the measurement |
Variables | process and the formulas. Input of the numerical values of the wavelengths, factors, constants, |
| etc. |
|
|
| Used to enter abbreviations and defined times for up to four timers. Highlight the appropriate line |
Timer 1, Timer 2, | and press Edit. The timers are activated or deactivated with the control boxes on the left of the |
Timer 3, Timer 4 | display. In the next column, a selection can be made from a list of names that describe the |
| corresponding work step. In the third column, the times for each active timer are entered. |
|
|
6.1.2.1 Measurement process
The measurement process defines the handling and the measurements of the test:
•At which and how many wavelengths should measurements be completed
•How many absorbance measurements must be completed?
•When should the zero measurement and the sample measurement be completed?
•Are waiting times necessary between measurements?
72