Quick Check® 600/800 Series User’s Guide 4 - 9
Reviewing the Dimensional Parameters
The following table lists the dimensional parameters and summarizes their
results. Descriptions of the parameters follow the tables.
* The method you select (Avg Decode or ISO/IECISO/IEC) determines the scan and sym-
bol grades.

Errors in Dimensional Parameters

If, after a scan, the abbreviation for dimensional appears as >DIM<, an error
exists in one or more of the dimensional parameters. If there is an error, use the
arrow buttons to review each parameter to identify the problem. If the
abbreviation for dimensional appears as Dim, the symbol has no dimensional
errors.

Average Bar

The average bar print error (AvgBar) value is the amount of average bar growth
or loss in the measured symbol relative to its “X” dimension. This value is
compared against traditional print tolerance criteria and rated. If the AvgBar is
within these limits, it is rated OK! or pass. If the AvgBar is outside these limits,
this test fails.
Parameter Trad’l ISO/IEC
Average Bar Error (Avg Bar Err)
Wide-to-narrow ratio (Wide/Nar Ratio)
Decodability*
Parameter Res ult
Avg Bar Err Compared against traditional print tolerance crite-
ria set by symbology or application
OK if within criteria
Fail if outside criteria
Wide/Narrow Ratio Compared against generic or user set limits for the
symbology or application
OK if within limits
Fail if outside limits
Decodability A if > .62
B if > .50
C if > .37
D if > .25
F if < .25