SECTION 8—BATTERIES

 

Connector on

Connector on

Battery Box

Battery Tray

(Hidden from View

 

Battery Box

 

Handle

 

Battery Box

Battery Tray

Wheelchair Base

FIGURE 8.3 Removing/Installing the Battery Box on the At’m QT

Removing/Installing the Batteries From/Into the At’m Battery Pack

NOTE: For this procedure, refer to FIGURE 8.4 on page 48.

Removing

1.Remove the battery pack from the wheelchair. Refer to Removing/Installing the Battery Pack on the At’m on page 45.

2.Open the battery pack by performing the following -

A.Grasp the side flaps and pull up.

B.Grasp the front flap and pull up.

3.Disconnect the wiring harness from the batteries in the following order -

A.NEGATIVE (-) BLACK battery cable from the NEGATIVE (-) battery terminal on the left battery.

B.POSITIVE (+) RED battery cable from POSITIVE (+) battery terminal/post on the right battery.

C.WHITE battery cable (jumper) from the POSITIVE (+) battery terminal/post on the left battery and the NEGATIVE (-) battery terminal/post on the right battery.

4.Remove the batteries from the battery pack.

Part No. 1125078

47

At’m and At’m QT

Page 47
Image 47
Invacare 1125078 service manual Removing/Installing the Battery Box on the At’m QT