Invacare Pronto M71 manual Comments, Information Gauge Display, Description

Models: Pronto M71

1 80
Download 80 pages 29.46 Kb
Page 26
Image 26
COMMENTS

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

Information Gauge Display Diagnostics

DISPLAY

DESCRIPTION

DEFINITION

COMMENTS

 InformationDESCRIPTIONManual backgroundManual backgroundManual backgroundManual backgroundManual background Information

Gauge

Display

All LEDs are off.

Power is off.

 

 

 

 

All LEDs are on.

Power is on.

Fewer than three LEDs on

 

 

 

 

 

 

implies reduced battery

 

 

 

 

 

 

charge.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Left RED LED is flashing.

Battery charge is low.

The batteries should be

 

 

 

 

 

 

charged as soon as possi-

 

 

 

 

 

 

ble.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Left to Right “chase” alter-

Joystick is in program-

The steady LEDs indicate

 

 

 

 

nating with steady display.

ming, inhibit and/or charg-

the current state of the

 

 

 

 

 

ing mode.

battery charge.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

All LEDs are flashing slowly.

Joystick has detected Out-

Release the joystick back

 

 

 

 

 

of-Neutral-at-Power-Up

to Neutral.

 

 

 

 

 

mode.

 

 

 

 

 

 

 

Service Indicator Light Diagnostics

NUMBER

ERROR CODE

 

OF

POSSIBLE SOLUTION

DESCRIPTION

FLASHES

 

 

 

 

 

 

1

User Fault

Release joystick to neutral and try again.

 

 

 

2

Battery Fault

Charge the batteries. Refer to Charging Batteries on page 74.

 

 

Check that battery cables are connected properly. Refer to

 

 

Connecting/Disconnecting the Battery Wiring Harness on page 72.

 

 

If necessary, replace batteries. Refer to Installing/Removing the

 

 

Batteries on page 70.

 

 

 

3

Left Motor Fault

Contact Invacare/Dealer for service.

 

 

 

4

Right Motor Fault

Contact Invacare/Dealer for service.

 

 

 

5

Left Park Brake Fault

Contact Invacare/Dealer for service.

 

 

 

6

Right Park Brake Fault

Contact Invacare/Dealer for service.

 

 

 

7

Remote Fault

Check to make sure joystick is connected properly.

 

 

Contact Invacare/Dealer for service.

 

 

 

8

Controller Fault

Contact Invacare/Dealer for service.

 

 

 

9

Communications Fault

Contact Invacare/Dealer for service.

 

 

 

10

General Fault

Contact Invacare/Dealer for service.

 

 

 

11

Incompatible or incorrect

Wrong type of remote connected. Contact Invacare/Dealer for

 

Remote

service.

 

 

 

 

 

 

Pronto® M94™

26

Part No. 1122145

Page 26
Image 26
Invacare Pronto M71 manual Comments, Information Gauge Display, Description