SECTION
Information Gauge Display Diagnostics
DISPLAY | DESCRIPTION | DEFINITION | COMMENTS |
Information
Gauge
Display
All LEDs are off. | Power is off. |
|
|
|
| All LEDs are on. | Power is on. | Fewer than three LEDs on |
|
|
|
|
|
| implies reduced battery |
|
|
|
|
|
| charge. |
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| Left RED LED is flashing. | Battery charge is low. | The batteries should be |
|
|
|
|
|
| charged as soon as possi- |
|
|
|
|
|
| ble. |
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| Left to Right “chase” alter- | Joystick is in program- | The steady LEDs indicate |
|
|
|
| nating with steady display. | ming, inhibit and/or charg- | the current state of the |
|
|
|
|
| ing mode. | battery charge. |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| All LEDs are flashing slowly. | Joystick has detected Out- | Release the joystick back |
|
|
|
|
| to Neutral. | |
|
|
|
|
| mode. |
|
|
|
|
|
|
|
Service Indicator Light Diagnostics
NUMBER | ERROR CODE |
| |
OF | POSSIBLE SOLUTION | ||
DESCRIPTION | |||
FLASHES |
| ||
|
| ||
|
|
| |
1 | User Fault | Release joystick to neutral and try again. | |
|
|
| |
2 | Battery Fault | Charge the batteries. Refer to Charging Batteries on page 74. | |
|
| Check that battery cables are connected properly. Refer to | |
|
| Connecting/Disconnecting the Battery Wiring Harness on page 72. | |
|
| If necessary, replace batteries. Refer to Installing/Removing the | |
|
| Batteries on page 70. | |
|
|
| |
3 | Left Motor Fault | Contact Invacare/Dealer for service. | |
|
|
| |
4 | Right Motor Fault | Contact Invacare/Dealer for service. | |
|
|
| |
5 | Left Park Brake Fault | Contact Invacare/Dealer for service. | |
|
|
| |
6 | Right Park Brake Fault | Contact Invacare/Dealer for service. | |
|
|
| |
7 | Remote Fault | Check to make sure joystick is connected properly. | |
|
| Contact Invacare/Dealer for service. | |
|
|
| |
8 | Controller Fault | Contact Invacare/Dealer for service. | |
|
|
| |
9 | Communications Fault | Contact Invacare/Dealer for service. | |
|
|
| |
10 | General Fault | Contact Invacare/Dealer for service. | |
|
|
| |
11 | Incompatible or incorrect | Wrong type of remote connected. Contact Invacare/Dealer for | |
| Remote | service. | |
|
|
| |
|
|
|
Pronto® M94™ | 26 | Part No. 1122145 |