Connecting the instrument to mains

4. Setting up the instrument

Connecting the instrument to mains

The instrument MUST be connected to an earthed mains power outlet socket.

It is recommended that the ASP300 be plugged into a wall socket that has Ground Fault Circuit Interruption (GFCI) protection - as an additional electrical safeguard.

Make sure to use the appropriate mains cable for the local volt- age supply (wall outlet).

Connect mains cable (6) to mains power supply socket (1) (see Fig. 3).

Jumper cable (3) connects the mains power supply output (2) to the electronics module input (4).

Switch ON/OFF switch (5) ON.

Switch ON/STOP switch at the right of the instrument ON.

 

Electrical

 

 

connections at the

 

1

rear panel of the

 

instrument

 

 

6

 

 

5

4

2

3

 

Cables

 

connected

Fig. 3

correctly

Important:!

Specification for connection (2) is as follows: 100 - 120 V or 230 - 240 V, max. 200 VA.

20

Instruction manual V 1.3 – 12/2003